Spontaneous locking of the knee after anterior cruciate ligament reconstruction as a result of a broken tibial fixation device
Tools
Metcalfe, Andrew J., James, Stuart H. and Fairclough, John A. (2008) Spontaneous locking of the knee after anterior cruciate ligament reconstruction as a result of a broken tibial fixation device. Arthroscopy: The Journal of Arthroscopic & Related Surgery, 24 (10). pp. 1195-1197. doi:10.1016/j.arthro.2007.05.005 ISSN 0749-8063.
Research output not available from this repository.
Request-a-Copy directly from author or use local Library Get it For Me service.
Official URL: http://dx.doi.org/10.1016/j.arthro.2007.05.005
Abstract
The Intrafix device (DePuy Mitek, Raynham, MA) is one of a number of recently developed products whose aim is to improve fixation of quadrupled hamstring grafts when used for anterior cruciate ligament reconstruction. We present a case of failure and intra-articular migration of the sleeve of an Intrafix device causing locking of the knee 10 weeks after anterior cruciate ligament reconstruction. We were unable to identify the cause of the failure or migration of the device. Rehabilitation had been progressing normally and without incident. The broken fragments were removed arthroscopically, and the reconstruction was found to be intact and healing well. Presumably, the device retained enough mechanical function to allow healing to progress, despite failure of the sleeve. This is, to our knowledge, the first reported case of such an event occurring with the new generation of hamstring graft fixation devices. © 2008 Arthroscopy Association of North America.
Item Type: | Journal Article | ||||
---|---|---|---|---|---|
Subjects: | R Medicine > R Medicine (General) | ||||
Divisions: | Faculty of Science, Engineering and Medicine > Medicine > Warwick Medical School > Clinical Trials Unit Faculty of Science, Engineering and Medicine > Medicine > Warwick Medical School > Health Sciences Faculty of Science, Engineering and Medicine > Medicine > Warwick Medical School |
||||
Journal or Publication Title: | Arthroscopy: The Journal of Arthroscopic & Related Surgery | ||||
Publisher: | Elsevier | ||||
ISSN: | 0749-8063 | ||||
Official Date: | 1 October 2008 | ||||
Dates: |
|
||||
Volume: | 24 | ||||
Number: | 10 | ||||
Page Range: | pp. 1195-1197 | ||||
DOI: | 10.1016/j.arthro.2007.05.005 | ||||
Status: | Peer Reviewed | ||||
Publication Status: | Published | ||||
Access rights to Published version: | Restricted or Subscription Access | ||||
Date of first compliant deposit: | 23 July 2018 |
Request changes or add full text files to a record
Repository staff actions (login required)
View Item |