Skip to content Skip to navigation
University of Warwick
  • Study
  • |
  • Research
  • |
  • Business
  • |
  • Alumni
  • |
  • News
  • |
  • About

University of Warwick
Publications service & WRAP

Highlight your research

  • WRAP
    • Home
    • Search WRAP
    • Browse by Warwick Author
    • Browse WRAP by Year
    • Browse WRAP by Subject
    • Browse WRAP by Department
    • Browse WRAP by Funder
    • Browse Theses by Department
  • Publications Service
    • Home
    • Search Publications Service
    • Browse by Warwick Author
    • Browse Publications service by Year
    • Browse Publications service by Subject
    • Browse Publications service by Department
    • Browse Publications service by Funder
  • Help & Advice
University of Warwick

The Library

  • Login
  • Admin

Clinical diagnostic accuracy of rapid detection of healthcare-associated bloodstream infection in critical care using multi-pathogen real-time polymerase chain reaction (RT-PCR) technology

Tools
- Tools
+ Tools

Warhurst, Geoffrey, Dunn, Graham, Chadwick, Paul, Blackwood, Bronagh, McAuley, Daniel, Perkins, Gavin D., McMullan, Ronan, Gates, Simon, Bentley, Andrew, Young, Duncan, Carlson, Gordon L and Dark, Paul (2015) Clinical diagnostic accuracy of rapid detection of healthcare-associated bloodstream infection in critical care using multi-pathogen real-time polymerase chain reaction (RT-PCR) technology. Health Technology Assessment, 19 (35). pp. 1-142. doi:10.3310/hta19350 ISSN 1366-5278.

Research output not available from this repository.

Request-a-Copy directly from author or use local Library Get it For Me service.

Official URL: http://dx.doi.org/10.3310/hta19350

Request Changes to record.

Abstract

Background
There is growing interest in the potential utility of real-time polymerase chain reaction (PCR) in diagnosing bloodstream infection by detecting pathogen deoxyribonucleic acid (DNA) in blood samples within a few hours. SeptiFast (Roche Diagnostics GmBH, Mannheim, Germany) is a multipathogen probe-based system targeting ribosomal DNA sequences of bacteria and fungi. It detects and identifies the commonest pathogens causing bloodstream infection. As background to this study, we report a systematic review of Phase III diagnostic accuracy studies of SeptiFast, which reveals uncertainty about its likely clinical utility based on widespread evidence of deficiencies in study design and reporting with a high risk of bias.

Objective
Determine the accuracy of SeptiFast real-time PCR for the detection of health-care-associated bloodstream infection, against standard microbiological culture.

Design
Prospective multicentre Phase III clinical diagnostic accuracy study using the standards for the reporting of diagnostic accuracy studies criteria.

Setting
Critical care departments within NHS hospitals in the north-west of England.

Participants
Adult patients requiring blood culture (BC) when developing new signs of systemic inflammation.

Main outcome measures
SeptiFast real-time PCR results at species/genus level compared with microbiological culture in association with independent adjudication of infection. Metrics of diagnostic accuracy were derived including sensitivity, specificity, likelihood ratios and predictive values, with their 95% confidence intervals (CIs). Latent class analysis was used to explore the diagnostic performance of culture as a reference standard.

Results
Of 1006 new patient episodes of systemic inflammation in 853 patients, 922 (92%) met the inclusion criteria and provided sufficient information for analysis. Index test assay failure occurred on 69 (7%) occasions. Adult patients had been exposed to a median of 8 days (interquartile range 4-16 days) of hospital care, had high levels of organ support activities and recent antibiotic exposure. SeptiFast real-time PCR, when compared with culture-proven bloodstream infection at species/genus level, had better specificity (85.8%, 95% CI 83.3% to 88.1%) than sensitivity (50%, 95% CI 39.1% to 60.8%). When compared with pooled diagnostic metrics derived from our systematic review, our clinical study revealed lower test accuracy of SeptiFast real-time PCR, mainly as a result of low diagnostic sensitivity. There was a low prevalence of BC-proven pathogens in these patients (9.2%, 95% CI 7.4% to 11.2%) such that the post-test probabilities of both a positive (26.3%, 95% CI 19.8% to 33.7%) and a negative SeptiFast test (5.6%, 95% CI 4.1% to 7.4%) indicate the potential limitations of this technology in the diagnosis of bloodstream infection. However, latent class analysis indicates that BC has a low sensitivity, questioning its relevance as a reference test in this setting. Using this analysis approach, the sensitivity of the SeptiFast test was low but also appeared significantly better than BC. Blood samples identified as positive by either culture or SeptiFast real-time PCR were associated with a high probability (>รข 95%) of infection, indicating higher diagnostic rule-in utility than was apparent using conventional analyses of diagnostic accuracy.

Conclusion
SeptiFast real-time PCR on blood samples may have rapid rule-in utility for the diagnosis of health-care-associated bloodstream infection but the lack of sensitivity is a significant limiting factor. Innovations aimed at improved diagnostic sensitivity of real-time PCR in this setting are urgently required. Future work recommendations include technology developments to improve the efficiency of pathogen DNA extraction and the capacity to detect a much broader range of pathogens and drug resistance genes and the application of new statistical approaches able to more reliably assess test performance in situation where the reference standard (e.g. blood culture in the setting of high antimicrobial use) is prone to error.

Study registration
The systematic review is registered as PROSPERO CRD42011001289.

Item Type: Journal Article
Divisions: Faculty of Science, Engineering and Medicine > Medicine > Warwick Medical School > Clinical Trials Unit
Faculty of Science, Engineering and Medicine > Medicine > Warwick Medical School > Health Sciences
Faculty of Science, Engineering and Medicine > Medicine > Warwick Medical School
Journal or Publication Title: Health Technology Assessment
Publisher: NIHR Health Technology Assessment programme
ISSN: 1366-5278
Official Date: May 2015
Dates:
DateEvent
May 2015Published
Volume: 19
Number: 35
Page Range: pp. 1-142
DOI: 10.3310/hta19350
Status: Peer Reviewed
Publication Status: Published
Access rights to Published version: Open Access (Creative Commons)

Request changes or add full text files to a record

Repository staff actions (login required)

View Item View Item
twitter

Email us: wrap@warwick.ac.uk
Contact Details
About Us